Tektronix P6042 Current Probe

Tektronix P6042 Current Probe 50MHz amplifier

Tektronix P6042 Current Probe 50 MHz

  • Bandwidth DC to 50 MHz

  • Aberrations <3% from 1mA/DIV to 50mA/DIV

  • Noise Periodic <0.5mA +0.2 div, tangentially measured

  • Random Trace Shift <1.5mA

  • Dynamic Range +500mV to -500mV at probe amplifier

  • Positioning Range +300mV to -300mV at probe amplifier output

  • Attenuator Accuracy within +/-3%

  • Required Output Load Impedance 50 Ohm

  • Input Power <20W

  • Risetime 7ns

  • Maximum Input Current 1A to 5mA: 10A (DC+peak AC), depending on frequency. 2mA and 1mA: 0.5A (DC+peak AC)

  • Thermal Drift (Probe only) <2mA/°C from 15°C to 35°C with amplifier temperature constant.

  • Thermal Drift (Amplifier only) <0.2div/°C with probe temperature constant.

  • Voltage-Feedthrough Susceptibility <250µA/V at 50MHz

  • External Magnetic Field Suscceptibility <250µA/Gauss, at 60Hz

Tektronix P6042 Stromzange 50MHz Stromzangenverstärker

Ein sehr wichtiges Instrument in jedem Elektronik Labor, eine Stromzange, die höhere Frequenzen und auch DC messen kann. Die Tektronix P6042 Stromzange wurde entwickelt in Verbindung mit einem 50 Ohm Eingang, der auf 50mV/DIV gestellt werden muss. Wer beispielsweise ein Schaltnetzteil entwickeln will, kommt ohne eine Stromzange eigentlich fast nicht mehr aus.

A very important Instrument in every laboratory, a current probe to measure from DC to higher frequencies. The Tektronix P6042 was designed for use with scopes having either 50 Ohm or high-impedance inputs. Set the scope to 50mV/DIV, terminate the coaxial cable with 50 Ohm. For somebody developing a switching power supply, a current probe  is a big help.

www.amplifier.cd        Tektronix Gallery        Index