The Model 6451 Databridge component measurement
instrument is an economically priced, automatic and microprocessor controlled
LCR bridge with a basic accuracy of 0.1%. Features:
|
|
|
|
Basic Accuracy
|
0.1% of
reading, ±1 digit
|
|
R
|
Measurement Range
|
0.1mΩ - 990MΩ
|
Ultimate Resolution
|
0.1mΩ
|
|
Conditions
for Basic Accuracy |
100
Hz/120Hz:- 2Ω - 1MΩ, 1 kHz:- 2Ω -500kΩ, 10kHz:- 2Ω - 100kΩ |
|
L
|
Measurement Range
|
0.001μH -9900H
|
Ultimate Resolution
|
0.001μH
|
|
Conditions
for Basic Accuracy |
100Hz:- 4mH
- 2000H, 1kHz:- 400μH - 200H, 10kHz:- 40μH - 10H |
|
C
|
Measurement Range
|
0.001pF -99mF
|
Ultimate Resolution
|
0.001pF
|
|
Conditions
for Basic Accuracy |
100Hz:- 4nF
- 2000μF, 1kHz:- 400pF - 200μF, 10kHz:- 40pf - 10μF |
|
Q
|
Measurement Range
|
0.001- 999 |
Ultimate Resolution
|
0.001
|
|
Conditions
for Basic Accuracy |
0.25 - 4
|
|
D
|
Measurement Range
|
0.001- 999
|
Ultimate Resolution
|
0.001
|
|
Conditions
for Basic Accuracy |
0.25 - 4
|
|
|
|
|
Maximal Voltage on Component
|
0.3V rms
|
|
DC-Bias Voltage
|
2V internal (selectable) or
up to 50V external
|
|
Measurement Modes
|
Series or parallel
equivalent circuit
|
|
Measurement Update Rate
|
2 per second
|
|
Temperature Range
|
0°C - 50°C
|
|
Power
|
95V to 125V or 195V to 255V,
48 - 63Hz
|
|
Dimensions
|
440 x 260 x 100 mm
|
|
Weight
|
6.5 kg |