Isolated Power Supply world-wide AC input

Part 7

Changing Clamping Diode TVS350V


062
Figure 140
Evaluation Board use TVS350 Clamping Diode.

(Current has a measurement error: current probe amplifier terminated with 50ohm scope input + 50ohm resistor additional on BNC cable, so real current is a higher)


Primary Coil Waveform

063
Figure 141
Primary-Power-Coil max. voltage +400Vpk/-300Vpk. Drain potential = GND. Transformer isolation must isolate this voltage and AC waveform, check datasheet.

Repeat test outside normal operating conditions, under increased supply voltage. Do these extended test to a time when you are familar with the circuit.

Don´t worry about the current peak on beginning of the rising ramp, caused by the 100:1 probe alligator clamp on Drain potential. Drain potential is very dynamic compared to the large surface oscilloscope ground.

(Current has a measurement error: current probe amplifier terminated with 50ohm scope input + 50ohm resistor additional on BNC cable, so real current is a higher)


064
Figure 142
Same measurement like in figure 141. The probe alligator clamp connected to the constant low impedance rectified DC signal. Probe tip connected to Drain.

Current Peak in the beginning of the rising ramp is much lower and looks ok. Drain signal sensitive to capacitive load, part of the LC-tank.

(Current has a measurement error: current probe amplifier terminated with 50ohm scope input + 50ohm resistor additional on BNC cable, so real current is a higher)


065
 Figure 143
Voltage on auxiliary primary coil, supplying the IC with power and information about the primary power coil voltage waveform. Primary auxiliary coil used for feedback regulation of the secondary side output voltage and finding the correct switching points.

Compare voltage waveforms of figure 142 with 143, looking similar - transformer works well.

The method of using two coupled primary coils is a advantage of this IC, prevents isolating coupling devices (e.g. optocoupler) between primary and secondary side. Reduce part count, simplifies layout and improve reliablility.

(Current has a measurement error: current probe amplifier terminated with 50ohm scope input + 50ohm resistor additional on BNC cable, so real current is a higher.)

IC Supply Voltage

PFC IC supply voltage
Figure 144
@85VAC rms, power supply Vcc bypassed by 47µF Electrolytic in parallel with 100nF ceramic.


PFC Ic supply voltage 260

Figure 145
@260VAC rms, power supply Vcc bypassed by 47µF Electrolytic in parallel with 100nF ceramic.





Figure 146
hot core @260VAC rms, 30W out, 28VDC out

Maximum possible output power 30 watt under 265VACrms, current limit 1.7Apk reached. Core saturated, running warm within 3 minutes. Current ramp very sharp, inductance decreased under high currents.

Trying to limit the Output Power

There was the idea to build a circuit around FB and ZCD - not used in this application.

066
Figure 147
FB feedback. ZCD, voltage on auxiliary winding.




067
Figure 148
@110VAC rms, 16.3Wout, 28VDCout


068
Figure 149
@150VAC rms, 16.3Wout, 28VDCout


069
Figure 150
@200VAC rms, 16.3Wout, 28VDCout


070
Figure 151
@250VAC rms, 16.3Wout, 28VDCout


071
Figure 152
@280VAC rms, 16.3Wout, 28VDCout

Comparison FB and ZCD

sweeping from @ 90 - 280VAC rms,                @16.3Wout, 28VDC out
90VAC 100VAC 110VAC 130VAC 150VAC 175VAC 200VAC 225VAC 250VAC 280VAC
FB max 3.664V 3.471V 3.392V 3.347V 3.323V 3,310V 3.319V 3,297V 3.288V 3.291V
FB peak 2.202V 1.948V 1.858V 1.837V 1.6V 1,426V 1.479V 1,513V 1.542V 1.599V
FB mean 2.8835V 2.7830V 2.688V 2.5368V 2.455V 2,3744V 2.321V 2,2701V 2.239V 2.192V
ZCD max 3.25V 3.25V 3.24V 3.22V 3.21V 3,19V 3.16V 3,12V 3.09V 3.07V
ZCD mean 0.914V 0.971V 1.02V 1.098V 1.166V 1,242V 1.298V 1,346V 1.384V 1.43V
Figure 153
Purpose of the table, getting a better understanding of Pin behaviour under changing AC voltage

FB ZCD vs line
Figure 154
Constant output power 16.3Wout 28VDC. Only the mean values are useful for a possible circuit.


sweeping from @ 90 - 280VAC rms,                @9.3Wout, 28.9VDC out
90VAC 100VAC 110VAC 130VAC 150VAC 175VAC 200VAC 225VAC 250VAC 280VAC
FB max 3.124V 3.085V 3.075V 3.088V 3.072V 3.059V 3.027V 3.038V 3.003V 3.007V
FB peak 1.498V 1.502V 1.527V 1.6V 1.635V 1.675V 1.68V 1.727V 1.723V 1.754V
FB mean 2.056V 2.016V 1.9908V 1.9447V 1.9043V 1.8646V 1.8282V 1.7957V 1.7654V 1.73626V
ZCD max 3.00V 2.97V 2.97V 2.93V 2.91V 2.88V 2.86V 2.85V 2.87V 2.91V
ZCD mean 0.892V 0.940V 0.981V 1.051V 1.11V 1.173V 1.222V 1.263V 1.303V 1.349V
Figure 155

9watt
Figure 156
Constant output power 9.3Wout 28.9VDC

9w_16W
Figure 157
9.3Wout and 16.3Wout in the same diagram


sweeping from @ 90 - 280VAC rms,                @Pmax. Wout at Switch-OFF,  28VDC out
90VAC 100VAC 110VAC 130VAC 150VAC 175VAC 200VAC 225VAC 250VAC 280VAC
FB max 3,749V 3,381V 3,360V 3,430V 3,475V 3,503V 3,571V 3,539V 3,553V 3,567V
FB peak 2,208V 1,326V 1,338V 1,397V 1,453V 1,524V 1,628V 1,581V 1,618V 1,669V
FB mean 2,9255V 3,0778V 3,0477V 3,0901V 3,1111V 3,1104V 3,1380V 3,1336V 3,1383V 3,1495V
ZCD max 3,25V 3,23V 3,25V 3,25V 3,25V 3,25V 3,26V 3,21V 3,22V 3,14V
ZCD mean 0,905V 0,969V 1,018V 1,106V 1,181V 1,262V 1,325V 1,356V 1,4V 1,412V
P max. 17,9W 20,37W 20,98W 22,06W 23,5W 25,15W 26,94W 28,92W 30,27W 30,86W
Load R 44,7 Ohm 39,6 Ohm 37,9 Ohm 35,8 Ohm 33,6 Ohm 31,4 Ohm 29,1 Ohm 27,1 Ohm 25,9 Ohm 25,4 Ohm
VDC out 28,3VDC 28,4VDC 28,2VDC 28,1VDC 28,1VDC 28,1VDC 28,0VDC 28,0VDC 28,0VDC 28,0VDC
Core PT1000 1181 Ohm 1185 Ohm 1182 Ohm 1189 Ohm 1201 Ohm 1235 Ohm 1241 Ohm 1236 Ohm 1235 Ohm 1240 Ohm
Temp 47,01°C 48,05°C 47,27°C 49,09°C 52,02°C 61,04°C 62,59°C 61,39°C 61,04°C 62,33°C
Figure 158
Over load condition under all AC-line-voltages

full load
Figure 159
Over load condition under all AC-line-voltages. Under these power levels controller stops switching. After cooling down transformer core, controller starts automatically. Controller manage core saturation very safe.


core temp
Figure 160
Highest core temperature under permanent max. overload condition.

Overload Condition means, operating with the lowest possible load resistance close to the Peak-Current-Switch-Off. Under Overload Condition the core becomes saturated, additional test under max. operating temperature recommended.

It was not possible to reach a higher core temperature than approx. 63°C

Controller manage core saturation absolute safe, no damaged parts. Evaluation Board works fine.


coretemp
Figure 161
Measurement of the core temperature.

PT1000 glued on the core. Covered with thermal heat-grease for a better thermal integration of the sensor. Main error source, convection on the small green wires. It is recommended keep wires close to the core.

Best measurement would be a sensor under the windings or more inside the core. Therefore the shown measurement can have 10-15°C measurement error compared to the inner core region. Compare measurement with a thermal camera or other methods. In this experiment it is not necessary to measure the max. temperature value with high precision, here the temperature amplitude vs. AC-voltage was a matter of interest.

Limit Output Power via external control

Experiments for external  power control
Messung CTRL
Figure 162
DC-supply on Control Pin, noisy scope plot


Messung Control
Figure 163
Pin has high impedance, collects all noise under open end wire

Messung Control
Figure 164
Measurement requires shielded coaxial cable.



Figure 165
It is possible to control output power via an external voltage, transfer function is nonlinear. Function not required, no further investigations done to implement a control circuit - end of experiment.


When doing some experiments with the FB pin, the Eval Board has been damaged by user mistake.

Controller and power supply zenerdiode renewed.

zener 15V curve
Figure 166
Current vs. Voltage curve of a working 15V zener diode


failed 16V zener curve
Figure 167
Current vs. Voltage curve of the damaged 16V zenerdiode. In forward direction curve looks ok, in reverse direction curve failed, compare with figure 166. In reverse direction failed diode behaves like a non-linear resistor, not like a zener.

Approx. R = -15V/-52mA ==> 288 Ohm

Don´t throw away such parts (but mark it). If you ever need for any experiment a non-linear resistor (1 quadrant only), use it. The question will be how stable remain this curve?



This failed zener curve was another good example for the need of special test equipment. No labority is complete without a curve-tracer, a highly recommended instrument.

Without such instruments it is more difficult to find failed parts, e.g. a DMM with a simple diode function tester would show an intact diode in forward direction !! the reverse direction the DMM would detect the failure.

But what happen if the user test only the zener forward direction with the DMM?, because user knows the DMM will not show a result for a 16V zener in reverse direction, no reverse test done. User would think this diode works correct and can be resoldered back on the PCB:

Such a scenario of undetectable errors (lack of test equipment, lack of experience, wrong assumptions), can made a user sick, e.g. wasting many hours searching for others (non-existing) errors, changing many parts on the PCB hoping always this was the fault - no success - the fault was already resoldered back on PCB by the user itself. At the end of the day, he will throw away the PCB and use a new PCB. This is sometimes a fast way, but not the way of understanding things and needs.



Figure 168
This was the experiment, external applied voltage on FB. Controller destroyed under experiments by user mistake. Evaluation Board works again after renewing controller and zener diode.

Go to Part 1
Go to Part 2
Go to Part 3
Go to Part 4
Go to Part 5
Go to Part 6
This is Part 7
Go to Part 8

Technische Berichte
www.amplifier.cd.

Impressum und Haftungsausschluss